Conas feabhas a chur ar thorthaí an 260/230 ríomh ag baint úsáide as an modh coirníní maighnéadach chun imoibrithe a bhaint as?

When extracting nucleic acids using the magnetic bead method, a low 260/230 cóimheas (de ghnáth < 1.8) indicates the presence of organic compounds(such as guanidine salts, feanól, eatánól, EDTA) or residual salts in the sample, which can affect downstream experiments (mar PCR, seicheamh).

 Here is a systematic optimization plan:

  1. Optimization of Extraction Steps

a. Complete sample lysis

Problem: Incomplete lysis leads to incomplete release of impurities.

Improvement: Increase the lysis time (e.g., extend to 30 nóiméad) or temperature (56℃).

Add a grinding step or additional proteinase K digestion for complex samples (such as plants, feces).

b. Thoroughly wash the magnetic beads

Key step: Incomplete washing is the main cause of 260/230 íseal.

Improvement: Increase the number of washes: 2 times → 3 amanna (especially for samples with high impurities).

Washing solution formula: Ensure to use freshly prepared 80% eatánól (avoid excessive water content or contamination).

Washing operation: Vortex or shake to mix the magnetic beads each time, avoiding the formation of beads clumps.

Thoroughly aspirate the supernatant (can briefly centrifuge to remove residual liquid).

c. Complete drying of magnetic beads

Problem: Residual ethanol can significantly reduce 260/230.

Improvement: Open the cap and let it dry for 5-10 nóiméad (at room temperature or 37°C), until there is no ethanol smell. Avoid excessive drying (otherwise, it will reduce the efficiency of nucleic acid elution).

 d. Optimize elution conditions

Use preheated elution buffer (such as 65°C TE buffer or sterile water), and let it stand for 5 minutes before collection.

Avoid using too small elution volume (recommend ≥ 50 μL to reduce salt concentration).

Ⅱ.Rialú Cáilíochta of Reagents and Consumables

a. Replace with fresh reagents:

Check if ethanol and washing solution have been contaminated (if stored for a long time without sealing).

Avoid using expired magnetic bead reagents (magnetic bead degradation may adsorb impurities).

b. Consumables selection:

Use centrifuge tubes and pipettes without nuclease or pyrogen (to avoid interference from plastic additives).

Ⅲ.Sample preprocessing

  1. Complex samples (such as soil and plants):

Increase the CTAB or SDS pretreatment to remove polysaccharides and polyphenols.

  • For blood samples:

 First remove hemoglobin using erythrocyte lysis solution.

  • Inhibitor-enriched samples (such as feces and body fluids):

Dilute the samples or use dedicated lysis buffers (such as those containing Carrier RNA).

Ⅳ.Verification and remedial measures

a. Electrophoresis or Qubit verification:

Check for degradation (tailing phenomenon) using agarose gel electrophoresis.

Compare the results with those from Nanodrop using a fluorescence meter (Qubit) to eliminate interference from impurity absorbance.

b. Purification remediation:

If the ratio of 260/230 remains low, purify the nucleic acid using a silica column or by ethanol precipitation again.

V. Turgnamh Rialaithe

For each extraction, a known standard sample (such as λDNA or cultured cells) is set up to confirm the stability of the reagents.

If the 260/230 ratio of the control sample is normal, the problem may lie in the sample being tested itself.

Common Problem Troubleshooting Checklist

PhenomenonPossible CauseRéiteach
260/230 is low but 260/280 is normalProtein or phenol contaminationStrengthen the washing and drying steps
Both 260/230 agus 260/280 are lowProtein or phenol contaminationOptimize the lysis, replace the lysis solution
Only the 230 peak is highOrganic solvent contaminationCheck the purity of ethanol or buffer

By following these steps, the ratio of 260/230 can be significantly increased (target > 1.8). If the problem persists, please feel free to contact Shanghai Lingjun Biotechnology at 18103799069.

Soláthraí

Shanghai Lingjun Biteicneolaíocht Co., Teo.bunaíodh i 2016 atá ina mhonaróir gairmiúil d'ábhair bithmhaighnéadacha agus imoibrithe eastóscadh aigéid núicléacha.

Tá taithí shaibhir againn in eastóscadh agus íonú aigéad núicléasach, íonú próitéine, scaradh cille, ceimileacanacht, agus réimsí teicniúla eile.

Úsáidtear ár gcuid táirgí go forleathan i go leor réimsí, mar thástáil leighis, tástáil ghéiniteach, taighde ollscoile, pórú géiniteach, agus mar sin de. Ní chuirimid ar fáil ach táirgí ach freisin is féidir tabhairt faoi OEM, ODM, agus riachtanais eile. Má tá riachtanas gaolmhar agat, bíodh leisce ort teagmháil a dhéanamh linn .

Nuashonruithe Nuachtlitir

Cuir isteach do sheoladh ríomhphoist thíos agus liostáil dár nuachtlitir