Dalam eksperimen biologi molekul, verifying the quality of nucleic acids extracted by magnetic bead method through agarose gel electrophoresis (gel berjalan) adalah satu langkah yang penting. Namun begitu, beginners often encounter some confusing electrophoresis phenomena. What information lies behind these abnormal phenomena? Then how can it be prevented?
1. Band tailing: A warning of nucleic acid degradation
When the band shows continuous smear-like tailing from top to bottom, it usually indicates that the nucleic acid has degraded. This might be because:
1. Degradation of the sample itself (failure to handle it promptly after collection or repeated freezing and thawing
2. Nuclease contamination is introduced during the operation process
3. The storage time after cracking is too long
4. Excessive mechanical shearing force (overly vigorous blowing and striking)
Preventive measures: Use fresh samples; Ensure that the experimental environment is clean and free from pollution; The operation process is gentle. After extraction is completed, test it promptly or store it at -80℃.
2. Band blurring: A signal indicating residual impurities
If the band diffusion is blurred and the boundary is not clear, the most common reason is:
1. Residual salt ions (incomplete washing
2. Protein or polysaccharide contamination
3. Ethanol residue (affecting electrophoretic mobility
Preventive measures: Ensure that all waste liquid is thoroughly sucked up after the last wash. Allow the magnetic beads to dry moderately (the surface should be dull but not cracked); For complex samples, the number of washes can be increased.
3. Smiling bands: Improper electrophoresis conditions
The two ends of the band curve upward in a “smile” shape, which is usually related to the electrophoresis conditions:
1. Excessive electric field intensity (too high voltage)
2. Uneven gel cooling
3. The ionic strength of the electrophoresis buffer is inappropriate or it has been reused too many times
Preventive measures: Appropriately reduce the electrophoresis voltage; Use freshly prepared electrophoresis buffer; Make sure the electrophoresis tank is equipped with an appropriate cooling device.
4. Other frequently asked questions
No band: It may be due to magnetic bead loss, insufficient elution or insufficient sample size
Abnormal bright spots: It may be RNA contamination (in DNA samples) or genomic DNA contamination (in RNA samples)
Professional advice: Positive and negative controls should be set up for each experiment. This helps to quickly determine whether the problem lies in the sample itself or the extraction process. Pada masa yang sama, remember that glue running is only the first step in quality inspection. For downstream experiments such as qPCR and sequencing, it is recommended to use Nanodrop or Qubit for precise quantification.
Understanding the causes and solutions of these common problems can not only help you achieve beautiful electrophoresis results, but also be the key to ensuring the success of downstream experiments. Good experimental habits and meticulous technical operations are the foundation for obtaining high-quality nucleic acid samples.
Pembekal
Shanghai Lingjun Biotechnology Co., Ltd.ditubuhkan pada 2016 yang merupakan pengilang profesional bahan biomagnet dan reagen pengekstrakan asid nukleik.
Kami mempunyai pengalaman yang kaya dalam pengekstrakan dan penulenan asid nukleik, penulenan protein, pemisahan sel, chemiluminescence, dan bidang teknikal yang lain.
Produk kami digunakan secara meluas dalam pelbagai bidang, seperti ujian perubatan, ujian genetik, penyelidikan universiti, pembiakan genetik, dan seterusnya. Kami bukan sahaja menyediakan produk tetapi juga boleh menjalankan OEM, ODM, dan keperluan lain. Jika anda mempunyai keperluan yang berkaitan, sila hubungi kami .